PRODUCT BULLETIN: ATTENTION i-STAT ANALYSER USERSBack2008-08-22
This Product Bulletin contains important information regarding the i-STAT analyser cartridge storage and handling recommendations. Please review the information carefully.
BACKGROUND: It has been determined that certain i-STAT analyser cartridges may exhibit an increased rate of quality check codes 45 and 55 when they are run without allowing sufficient time prior to use for the refrigerated cartridges to warm to room temperature. Your analyser may require service in order to become operational after the occurrence of quality check codes 45 and 55. Following the manufacturer's recommendations for cartridge warming will RECOMMENDED ACTIONS: 1. For long-term storage, keep cartridges in the refrigerator at 2-8° C. 2. Individual cartridges removed from the refrigerator should stand at room temperature for at least five (5) minutes prior to use. 3. Boxes of cartridges removed from the refrigerator should stand at room temperature for one (1) hour prior to use. 4. Cartridges that have been equilibrated to room temperature must not be returned to the refrigerator. 5. Once cartridges have been equilibrated to room temperature they may be stored